Keysight Technologies N2791A Oscilloscope Probe, Probe Type: Differential 25MHz 700V 1:10, 1:100

Technical Reference
Legislation and Compliance
RoHS Certificate of Compliance
COO (Country of Origin): MY
Product Details

Keysight High-voltage Differential Active Probes N2790A & N2791A

These high-voltage differential probes are for making, safely and accurately, floating measurements in power electronics circuits. They enable conventional earth-grounded Keysight oscilloscopes to be used for floating signal measurements (i.e. where neither point of measurement is at earth ground). Selectable attenuation settings make for highly versatile probes which come with tip accessories to facilitate working with large and small components in tight spaces.

Note

The N2790A has a differential amplifier in the probe head and an overrange indicator which alerts the user when the probe input exceeds the dynamic range of the probe. It is compatible with InfiniiVision 5000, 6000 (except 100MHz) and 7000 Series (with version 5.26.0001 software), Infiniium 8000, 54830 Series (with version 5.7 software) and Infiniium 9000 Series (with version 2.0 software). Each of these scopes can support four N2790A probes.

Keysight Technologies Oscilloscope Probe

The Keysight Technologies N2791A oscilloscope probe offers conventional earth-grounded Keysight oscilloscopes to be used for floating signal measurements of up to 700V of differential voltage and common-mode voltage. It offers sufficient dynamic range and bandwidth for your application to make the floating measurements found in power electronics circuits safely and accurately. It also has an over-range indicator which alerts the user when the probe input exceeds the dynamic range of the probe. It is compatible with InfiniiVision 5000, 6000 (except 100MHz) and 7000 Series (with version 5.26.0001 software), Infiniium 8000, 54830 Series (with version 5.7 software) and Infiniium 9000 Series (with version 2.0 software). Each of these scopes can support four N2790A probes.
Supplied with alligator clips (2), hook clips (2), USB power cord 2m, 4 AA batteries, manual.

Features and Benefits

• Bandwidth is DC to 25MHz (–3dB)
• Attenuation ratio is 10:1/100:1
• Input impedance is 4Mohm ll10pF each side to ground
• Differential range is (10:1) ±70V (DC + peak AC)
• Maximum differential is (100:1) ±700V (DC + peak AC)
• Common mode range is ±700V (100:1) (DC + peak AC) or 70V (10:1)
• Absolute maximum voltage is 1000VRMS CAT III (1/100 & 1/1000)
• Offset (typical) is ±7.5mV
• Source input impedance is 1Mohm BNC interface
• AC CMRR (typical) is -80dB at 50/60Hz and -40dB at 1MHz
• Compatible with any oscilloscope with 1 MΩ BNC input
• Ambient operating temperature ranges between -10°C and 50°C
• Power requirement 4xAA cells

Certifications

• IEC61010-031
• CAT II 1000 VAC

Supplied with

alligator clips (2), hook clips (2), USB power cord 2m, 4 AA batteries, manual

Specifications
Attribute Value
Probe Type Differential
Attenuation 1:10, 1:100
Bandwidth 25MHz
Maximum Common Mode Input Voltage 700V
Connector Type BNC
Model Number p N2791A
Safety Category Level CAT II
Safety Category Voltage 1000V ac
Series N
Safety Category CAT II 1000 V ac
Rise Time 14ns
10 In stock for FREE next working day delivery
Price Each
£ 594.00
(exc. VAT)
£ 712.80
(inc. VAT)
Units
Per unit
1 +
£594.00
Related Products
The RS PRO RS-HV series passive oscilloscope probe ...
Description:
The RS PRO RS-HV series passive oscilloscope probe is ergonomically designed and is suitable for use with all the major brands of oscilloscopes as well as our range. This probe does not require a power supply or batteries. Hence, it is lightweight and easily portable. It comes with a fixed ...
The HVD3000A series of high voltage differential probes ...
Description:
The HVD3000A series of high voltage differential probes cover the fullest range of applications, from 120/240 V switch-mode power supplies through 600 V class and 5 kV electrical apparatus. The table below provides a list of recommended probe models based on specific device under test applications. 1 kV, 2 kV, ...
When you use the oscilloscope in its usual ...
Description:
When you use the oscilloscope in its usual context to show and measure circuit shaft formations, you are working in the most common mode for voltage measurements. The differential probe enhances the oscilloscope probe's measuring capabilities with performance conversion, inverters, motor speed indicators, switch mode power supplies and many other ...
For high-speed switch mode power supply design; CAN/LIN ...
Description:
For high-speed switch mode power supply design; CAN/LIN bus design; high-speed digital design; digital design and characterization; manufacturing engineering test; research and development. ±42 V (DC + pk AC) differential input voltage 18 dB CMRR (at 250 MHz 50X attenuation) Selectable bandwidth limiting filters DC reject Connects to some Tektronix ...