This certificate confirms that the product detailed below complies with the specifications currently published by RS Components and has been subject to the quality conditions of our registration to the BS EN ISO9001:2008 management standard. Furthermore, where applicable, it provides assurance that all electrostatic discharge sensitive devices have been handled and packed under conditions that meet the administrative and technical requirements of the ANSI/ESD S20.20:2007 and the BS EN 61340-5-1:2007 electrostatic control standards.
RS Stock No.
146-777
Description
Kane 450 Gas analyser, CO, CO2, O2
Manufacturer/Brand:
Kane
Mfr Part No.
KANE450 KIT/RS
The foregoing information relates to product sold on, or after, the date shown below.
RS COMPONENTS
Date
Feb 9, 2012
RS Components Ltd, Birchington Road, Corby, Northamptonshire, NN17 9RS, UK
Specifications
Battery Life
>12h
Battery Life - Toxic Gas/Oxygen Models
>12
CO Measurement Accuracy
±0.2%
Display Type
Backlit Display
Gases Measured
O2 , CO2, CO & CO/CO2 Ratio
Still not found what you are looking for?
Not found what you are looking for? You can find products which may be more suitable by simply checking the features that interest you above and clicking the Find button. Find
Alternative for Kane 450 Gas analyser, CO, CO2, O2
The Kane 450 has an easy to use rotary switch and includes all the features expected of a top of the range analyser. Any four parameters can be displayed simultaneously. The high accuracy and stability of its pressure readings allow difficult tasks such as measuring flue draught and checking the air/fuel ratio valve on high efficiency boilers can be performed with confidence. It is fitted with a 0–4,000ppm CO sensor and is suitable for commercial and domestic boilers.
Four line backlit display controlled by rotary switch Measures O2, CO, temperature and pressure Calculates CO2, CO/CO2 ratio, excess air and efficiency Multi-fuel: natural gas, propane, butane, LPG, 28//35 sec oils, wood pellets Fitted with 0 – 4,000ppm CO sensor, hydrogen compensated High accuracy differential manometer Differential thermometer for flow / return / ΔT Hold, store and auto-store readings Reports include combustion test, tightness test, CO build-up and differential temperature Infra-red output to thermal paper printer Torch light Connection for optional gas leak detector Designed to meet BS7927, EN 50379 and BS7967 Battery life typically 12+ hours from full charge
Parameter
Resolution
Accuracy
Range
Temp Measurement
Flue Temperature
0.1°C/F
+2.0°C +0.3% reading
0-600°C
Inlet Temperature (Internal sensor)
0.1°C/F
+1.0°C +0.3% reading
0-50°C
Inlet Temperature (External sensor)
0.1°C/F
+2.0°C +0.3% reading
0-600°C
Gas Measurement
Oxygen
0.1%
±0.2%*1
0-21%
Carbon Monoxide
1ppm
±10ppm <100ppm ±5% reading <100ppm
0-4,000ppm
Carbon Dioxide *2
0.1%
±0.3% reading
0-30%
Efficiency *2
0.1%
±1.0% reading
0-99.9%
Excess Air *2
0.1%
±0.2% reading
0-250%
CO/CO2 ratio *2
0.0001
±5% reading
0-0.999
Pressure (differential) Nominal range +80mBar (Maximum over range without damage to sensor is +400mBar)